Tauc Plot Calculator
UV–Vis analysis

Band gap from UV-Vis data

How an optical band gap is extracted from a UV–Vis absorption or transmittance spectrum — the conversions, the Tauc plot and the checks that make the number trustworthy.

Absorbance · Transmittance · α

UV–Vis spectroscopy measures how much light a sample absorbs across the spectrum. From that single measurement you can estimate the optical band gap: convert the raw column into absorption, convert the wavelength axis into photon energy, and run aTauc ploton the rising edge.

The route depends on what your instrument exported — absorbance, percentage transmittance, or a pre-computed absorption coefficient. All three converge on the same fit, and theUV-Vis band gap calculatorhandles each conversion automatically.

Three data routes, one result

A → α → Tauc

From absorbance data

  1. Confirm the file is absorbance (not %T).
  2. Convert wavelength to photon energy: E = 1239.841984 / λ.
  3. If thickness d is known: α = 2.302585 · A / d.
  4. Plot (αhν)ⁿ vs hν (or (Ahν)ⁿ as the proxy).
  5. Fit the linear onset and extrapolate to Eg.
%T → A → α → Tauc

From transmittance data

  1. Convert percentage to fractional transmittance: T = %T / 100.
  2. Convert to absorbance: A = −log₁₀(T).
  3. Convert wavelength to photon energy.
  4. Apply thickness if available to obtain α.
  5. Fit the Tauc plot and read the intercept.
α → Tauc

From a pre-computed α file

  1. Confirm the column is α in cm⁻¹.
  2. Verify the energy axis is in eV.
  3. Apply the exponent n for your transition.
  4. Plot (αhν)ⁿ vs hν directly.
  5. Fit and extrapolate as usual.
A = −log10(%T / 100)
E (eV) = 1239.841984 / λ (nm)
α (cm−1) = 2.302585 · A / d (cm)
Eg = −b / m  from the fitted Tauc line

UV–Vis pitfalls that move the band gap

Scattering baseline

Powders and rough films scatter light, lifting the absorbance below the edge. A rising baseline inflates α in the tail and can bias the fit.

Detector saturation

Above roughly A = 2–2.5 most spectrophotometers flatten out. Exclude that plateau from the fit window.

Thin-film interference

Fabry–Pérot fringes superimpose oscillations on the edge. Choose a fringe-free region or average before fitting.

Wrong unit mode

Feeding %T into an absorbance-based workflow inverts the edge entirely. Always check the first rows of the file.

Instrument export quirks (what to check before you convert)

The step-by-step conversions are the same for every spectrophotometer — but the file layout is not. These are the export patterns that most often break a first-time Tauc analysis. The calculator’s delimiter and header detection covers the common layouts; still verify the column roles yourself once per instrument.

Shimadzu UV-Vis

TXT exports often mix a multi-line header with wavelength and Abs columns separated by tabs or multiple spaces. The calculator strips header rows automatically; confirm column 1 is nm (not cm⁻¹) before converting to eV.

PerkinElmer Lambda

CSV outputs may include an “Absorbance” or “Transmittance %” label row and a separate wavelength column. If the file says T%, declare transmittance mode — do not paste the percentage column into an absorbance workflow.

Agilent Cary

Cary WinUV exports can interleave baseline scans with sample scans. Keep only the sample spectrum, or subtract the baseline in the instrument software before the Tauc fit so the sub-gap region sits near zero.

Ocean Optics / OceanDirect

Process files are often two-column ASCII with wavelength in nm and intensity or absorbance depending on the processing step. If intensity is raw counts, convert to absorbance or transmittance first — raw counts are not a Tauc input.

For the full seven-step calculation (including exponent choice and validation), seehow to calculate band gap using a Tauc plot. This page stays focused on the UV–Vis data path itself.

Reading a UV–Vis spectrum before you fit

  • Does the low-energy region sit near zero absorbance? If not, baseline-correct first.
  • Is there a single edge, or two overlapping steps (mixed phases, defects)?
  • Where does the signal start flattening? That is your upper fit boundary.
  • Are there interference fringes on the film? Fit between them or average them out.
FAQ

UV-Vis band gap questions

Should I use absorbance or transmittance as the raw input?

Either, as long as you declare it correctly. Transmittance must be converted to absorbance first; absorbance can go straight into (Ahν)ⁿ or be converted to α when thickness is known. Mixed-up unit modes invert the edge.

Why does my absorbance spectrum not start at zero?

Scattering, a sloping baseline or solvent background lifts the sub-bandgap region. Subtract a baseline before the Tauc fit — otherwise the linear onset tilts and the extrapolated Eg shifts.

What should I check before trusting a UV-Vis band gap?

Confirm the unit mode (%T vs A vs α), convert wavelength to eV, exclude Urbach-tail and saturation points, pick n from the material physics, and verify that a small change in the fit window barely moves Eg.

Do interference fringes ruin a thin-film Tauc plot?

Fabry–Pérot oscillations superimposed on the edge distort linearity. Fit a fringe-free region, average over a fringe period, or use a film thick enough that fringes are washed out before extracting Eg.

Can I compare band gaps from different UV-Vis instruments?

Yes, if both are converted to the same axis (eV), the same unit mode and the same Tauc exponent, with comparable baselines. Absolute absorbance scale cancels in the intercept only when path length is consistent.

How does a double-beam blank affect the Tauc result?

A proper solvent/cuvette blank removes instrument offset. If the blank is wrong, the whole spectrum shifts vertically; the intercept can survive a pure scale error, but a sloping blank error will bias Eg.

Drop in your UV–Vis export

TXT, DAT and CSV files from common spectrophotometers are parsed automatically — absorbance or transmittance, wavelength or energy.